Homepage Analytical Instruments

Jenway PFP7/C Clinical Flame Photometer

Jenway PFP7 Flame Photometer

  • Na & K Filters as standard
  • Operates on Propane, Butane, Natural Gas or LPG supplies
  • Ca, Ba and Li Filters as optional accessories
  • Recorder Output
  • Electronic Flame Failure detection
  • Easily Maintained
  • Fine and Coarse Sensitivity controls
  • Economically priced

The Jenway Model PFP7/C is a low temperature, single channel Flame Photometer, designed for the routine determinations of Sodium (Na) and Potassium (K). Additional filters are available for the measurement of Lithium (Li), Barium (Ba) and Calcium (Ca).

The instrument is fitted with automatic flame failure detection for user safety, making it ideal for use in clinical, industrial and educational applications.

The Model PFP7/C is designed specifically for use in Clinical applications. The in-built lineariser circuitry enables readings of both Sodium and Potassium, at the normal clinical serum concentrations, to be directly displayed in mmol/l. Serum samples need to be diluted 200:1 or 100:1 prior to presentation to the flame photometer. To enable this to be carried out efficiently Jenway are able to offer the Model 8225 Diluter.

There is also an industrial version of this instrument, the Model PFP7.

SPECIFICATIONS
Ranges
120 - 160mmol/l Na (linearised) / 0 -10.0mmol/l K
Limits of Detection
Li <=0.25ppm / Ca <=15ppm / Ba <=30ppm
Reproducibility
<=1% / Coefficient of Variation for 20 consecutive samples using 10ppm Na set to read 50
Linearity
Better than 2% when concentration of 3ppm Na and K and 5ppm Li are set to read 100
Specificity
Interference from Na, K and Li equal in concentration to test element will be >0.5%
Recorder output
Nominal 1.00V for a reading of 100.0
Services: Electrical
90-125V or 190-250V @ 50/60Hz
Services: Air
Moisture and oil free 6 litre/minute @ 1Kg/cm2 (14psi)
Services: Fuel
Propane, Butane, Natural Gas or LPG
Size
420 x 360 x 300mm
Weight
8Kgs

ORDERING INFORMATION

500 200
Model PFP7/C supplied with Na & K filters, connecting hoses and clips, compressor plug, drain trap, fuses, mains lead and combined Instruction/Service manual

A number of additional accessories are available for the Flame Photometer range, including additional filters / regulators and a selection of clinical and industrial standard solutions.

Return to Top of Page